Analytical method study for the determination of radium-226 in water by liquid scintillation counting

GUO Chen, YANG Hailan, YANG Youkun, BAO Li, GOU Rongxia, WANG Ruijun, LIAN Bing   

  1. China Institute for Radiation Protection, Key Laboratory of Radiation Environment & Health of the Ministry of Ecology and Environment, Taiyuan 030006
  • Received:2023-11-23 Online:2024-11-20 Published:2024-12-26

摘要: 本文研究了水中镭-226的液体闪烁计数器测量方法,通过非水溶性闪烁液将镭-226的子体氡-222从水样中萃取进入闪烁液中,实现氡-222与水中其他干扰核素的分离。当镭-226与它的子体达到平衡后,通过液体闪烁计数器进行α/β甄别测量,测量氡-222、钋-218、钋-214的总α计数,间接得到水相中镭-226的活度浓度。结果表示:液闪对水中镭-226测量的淬灭很小,Quantulus 1220型液体闪烁计数器对本方法的测量本底为0.02 min-1,测量效率在2.52~2.60之间,探测下限小于2 mBq/L,可以满足实际环境样品中镭的测量需求。

关键词: 液闪, 镭-226, α/β甄别测量

Abstract: This paper studies a liquid scintillation counter measurement method for radium-226 in water. The method extracts radon-222, a daughter of radium-226, from water sample and enters into scintillation solution by means of insoluble scintillation liquid, thus realizes the separation of radon-222 from other interfering nuclides in water. After a period of time, when radium-226 and its daughter reached equilibrium, alpha/beta discrimination measurement was carried out by liquid scintillation counter to separate radon-222 and its alpha emmision daughters polonium-218 and polonium-214 from the beta emmision daughters bismuth-214 and lead-210, and indirectly obtains the activity concentration of radium-226 in the aqueous phase through the measurement of total alpha counts of radon-222, polonium-218 and polonium-214. The results indicated that the quench of liquid scintillation on the measurement of radium-226 in water is very low, and the Quantulus 1220 liquid scintillation counter has a measurement background of 0.02 cpm for this method, with a measurement efficiency of 2.52-2.60 and a minimum detection level of less than 2 mBq/L, which can meet the requirement for the measurement of radium in environmental samples.

Key words: LSC, radium-226, alpha/beta discrimination measurement