A compact LaBr3:Ce gamma spectrometer based on SiPM with high energy resolution

SHI Boxuan1, LIU Liye2, CAO Qinjian2, XIA Sanqiang2, WANG Xiaolong2   

  1. 1. CNNC Environmental protection Engineering Co. Ltd. , Beijing 100191;
    2. China Institute for Radiation Protection, Taiyuan 030006
  • Received:2019-09-03 Published:2020-07-10

摘要: 基于光电倍增管(photomultiplier tube,简称PMT)的LaBr3:Ce γ谱仪具有比NaI(Tl)γ谱仪更高的能量分辨率,但具有体积大、对磁场敏感、需要高电压等缺点。硅光电倍增管(Silicon photomultiplier tube,简称SiPM)具有与PMT相近的增益和效率,同时具有诸如高定时分辨率、抗磁场能力强、低偏压和紧凑尺寸等优良特性。本文将LaBr3:Ce晶体与SiPM阵列耦合,设计研制基于SiPM的紧凑型LaBr3:Ce γ谱仪,通过降噪、优化工作电压等措施改善SiPM的缺点对γ谱仪性能的影响。工作电压的噪声会导致能量分辨率发生恶化,通过设计无源滤波电路CLC π型滤波器,利用其对直/交流阻抗的不同特性,滤除高频纹波,工作电压的信噪比从未降噪前的62.6 dB提高到74.64 dB;能量分辨率最优值对应于表示暗噪声、串扰、光电探测效率和SiPM增益之间折衷的最佳工作电压。通过实验给出不同工作电压下的能量分辨率,确定最佳工作电压为54.8 V,该电压下的能量分辨率为3.06%(@662 keV),结果与使用光电倍增管(PMT)测量的2.89%非常接近。

关键词: γ谱仪, LaBr3:Ce晶体, SiPM

Abstract: The PMT-based LaBr3:Ce gamma spectrometer has higher energy resolution than NaI gamma spectrometer, but it has the disadvantages such as large volume, sensitive to magnetic field and high voltage demand. Silicon photomultiplier tube (SiPM) has similar gain and efficiency as PMT, while offering excellent characteristics such as high timing resolution, high magnetic field resistance, low bias and compact size. In this paper, the LaBr3:Ce crystal is coupled with the SiPM array, and the compact LaBr3:Ce γ spectrometer based on SiPM is designed and developed to improve the impact of the shortcomings of SiPM on the performance of gamma spectrometer by reducing noise, optimizing the operating voltage.The noise of working voltage will lead to the deterioration of energy resolution. By designing the CLC-type π filter with passive filter circuit, the high frequency ripple can be filtered by using its different characteristics of DC/AC impedance and the signal-to-noise ratio of working voltage can be increased from 62.6 dB to 74.64 dB before the noise reduction. The minimum energy resolution corresponds to the optimum operating voltage representing a compromise between dark noise, crosstalk, photodetection efficiency, and SiPM gain. The energy resolution at different working voltages is given by experiment, and the optimal working voltage is determined to be 54.8 V. The energy resolution at this voltage is 3.06% (@662 keV), the result of which is very close to 2.89% measured by photomultiplier tube (PMT).

Key words: gamma spectrometer, LaBr3:Ce crystal, SiPM