In-situ evaluation for activated concrete in accelerator facility with scintillation-type gamma-ray spectrometer

Go Yoshida1, Hiroshi Matsumura2, Koichi Nishikawa1, Akihiro Toyoda1, Yoshiharu Miyazaki2, Kazuyoshi Masumoto1, Hajime Nakamura1, Taichi Miura1   

  • 出版日期:2020-11-20 发布日期:2021-01-27

In-situ evaluation for activated concrete in accelerator facility with scintillation-type gamma-ray spectrometer

Go Yoshida1, Hiroshi Matsumura2, Koichi Nishikawa1, Akihiro Toyoda1, Yoshiharu Miyazaki2, Kazuyoshi Masumoto1, Hajime Nakamura1, Taichi Miura1   

  1. 1. High Energy Accelerator Research Organization KEK,1-1 Oho,Tsukuba,Ibaraki 305-0801,Japan;
    2. The Medical and Pharmacological Research Center Foundation,Wo-32,Inoyama-machi,Hakui,Ishikawa 925-0613,Japan
  • Online:2020-11-20 Published:2021-01-27
  • About author:The presenting author is an assistant professor at Radiation Science Center, KEK in Japan. His specialty is radiochemistry. He completed his Ph.D. in 2017 at Osaka Univ., Japan. E-mail:yoshigo@post.kek.jp

Abstract: The assessment of activated concrete is particularly difficult during the decommissioning of an accelerator facility.Destructive analysis by core boring is the only method of investigating the activity of concrete material.To address this problem,an in-situ and nondestructive analysis method was developed to determine γ-ray-emitting nuclides and their specific activities in the concrete walls and floor by using a portable germanium semiconductor detector.In this work,we examined a substitute for Ge detector to establish a simpler and more convenient method.As candidates,we focused on some scintillation type spectrometers,and the possibility of a substitute for a Ge detector was examined by both simulation and experiment.The detection limits were roughly estimated through Monte Carlo simulation for various scintillation crystals,and it was found that 1.5-inch LaBr3,CeBr3,and SrI2 could distinguish the clearance level.It was confirmed that the 1.5-inch LaBr3 could reproduce the calibration curve of the Ge detector in the experiment.The required thickness and length of the radiation shield for suppressing the background radiation during the measurement was also determined for the convenience of an actual decommissioning work.

Key words: activation, accelerator, concrete, scintillation detector, PHITS code