RADIATION PROTECTION ›› 2020, Vol. 40 ›› Issue (4): 271-277.

Previous Articles     Next Articles

Establishment of tritium activity concentration standard device

FENG Mei1, WEI Yingjing1, LI Dehong2, WANG Yong1, ZHANG Qingli1, WANG Mingliang1   

  1. 1.China Institute for Radiation Protection, Taiyuan 030006;
    2.National Institute of Metrology, Beijing 100013
  • Received:2019-05-14 Online:2020-07-20 Published:2020-09-21

Abstract: In order to solve the problem of the value transfer of tritium-in-air monitors in China,tritium activity concentration standard device was built up and the performance of the standard device was tested in laboratory. Testing results show that the device has a repeatability of 0.97%, a stability of 1.1%, activity concentration ranges between 5.0×104 Bq·m-3 to 2.0×108 Bq·m-3, and the relative extended uncertainty of 4.6%-5.0% (k=2) was achieved. This tritium activity concentration standard device has been authorized as social public measurement standard, which could be used for value transferring of tritium-in-air monitors in China.

Key words: tritium, activity concentration, tritium-in-air monitor, standard device, value transfer

CLC Number: 

  • TL84文