RADIATION PROTECTION ›› 2016, Vol. 36 ›› Issue (6): 364-367.

Previous Articles     Next Articles

Size adjustment and calculation of full-energy peak efficiency for the LaBr3 scintillation detector

Cao Qinjian, Xia Sanqiang, Liu Liye, Xiao Yunshi, Wang Yu, Li Hua   

  1. 1. China Institute of Atomic Energy,Beijing 102413;
    2. China Institute for Radiation Protection,Taiyuan 030006
  • Received:2015-07-15 Online:2016-11-20 Published:2021-11-15

Abstract: The technique of virtual calibration has been widely applied to field of calibration of the radiation detector. In this paper, the dimensions of the LaBr3 detector were adjusted based on the point-source experiments with Monte Carlo method. A series of full-energy peak efficiency of the detector were then simulated in different positions. The results of the validation experiment showed that the relative deviations between calculated efficiencies and measurement values are within ±5% in the energy range of 121—1 332 keV.

Key words: LaBr3 scintillation detector, Monte Carlo method, crystal adjustment, efficiency calibration

CLC Number: 

  • R144.1