The impact of additional Cu filtration on radiation dose in digital chest radiography

MIN Gaoyu, LI Xin, ZHAO Feng, HE Wei, HU Yibin   

  1. Clinical Engineering Department, The First Affiliated Hospital of Nanjing Medical University, Nanjing 210029
  • Received:2021-07-26 Published:2022-07-01

摘要: 采用西门子公司生产的数字化X射线透视摄影系统,在不同的附加铜滤过(0、0.1、0.2、0.3 mm)下,对不同厚度的聚甲基丙烯酸甲酯(PMMA)模体进行胸部摄影。利用对比度噪声比评估不同附加铜滤过对图像质量的影响,同时使用辐射剂量仪记录不同铜滤过下的体表入射剂量,以探讨数字化胸部X射线摄影中,附加铜滤过对受检者体表入射剂量的影响。结果表明,不同附加铜滤过下,图像的对比度噪声比没有明显变化。在不同厚度的PMMA模体下,附加0.1 mm铜、0.2 mm铜、0.3 mm铜后的体表入射剂量,与没有添加铜滤过时相比分别下降了24%~29%、33%~42%和39%~47%,且差别均有统计学意义(p<0.05)。说明在数字化胸部X射线摄影中,适当增加铜滤过,不会影响图像质量,但能有效减少受检者的体表入射剂量。

关键词: 铜滤过, 数字化胸部X射线摄影, 体表入射剂量, 对比度噪声比

Abstract: By using the digital X ray fluoroscopy camera system produced by Siemens, under different additional copper filters (0, 0.1, 0.2, 0.3 mm), chest photography is performed on PMMA phantoms of different thicknesses. Using the contrast-to-noise ratio to evaluate the impact of different additional copper filters on image quality, and using a radiation dosimeter to record the entrance surface dose under different copper filters, in order to explore the effects of additional copper filters on the body surface dose in digital chest X ray photography. The results show that under different additional copper filters, the contrast-to-noise ratio of the image does not change significantly. Under the PMMA phantoms with different thicknesses, the entrance surface dose decreased by 24%-29%, 33%-42% and 39%-47%, after adding 0.1 mm copper, 0.2 mm copper, and 0.3 mm copper respectively, compared with the absence of copper filtration, and the difference was statistically significant (p<0.05). It shows that for the digital chest X ray photography, the appropriate increase of copper filtering will not affect the image quality, but it can effectively reduce the body surface dose.

Key words: Cu filter, digital chest radiography, entrance surface dose, contrast to noise ratio