辐射防护 ›› 2021, Vol. 41 ›› Issue (S1): 145-150.

• 医疗照射与防护 • 上一篇    

γ射线空气比释动能(治疗水平)标准装置的建立

李蔚铭1, 韦应靖1,2, 李胤1, 陈双强1, 王明亮1, 郝世东1   

  1. 1.中国辐射防护研究院,太原 030006;
    2.清华大学,北京 100084
  • 收稿日期:2021-01-14 出版日期:2021-10-25 发布日期:2021-11-26
  • 通讯作者: 韦应靖。E-mail:weiyj531@sina.com
  • 作者简介:李蔚铭(1997—),男,2019年毕业于四川大学核工程与核技术专业,现为中国辐射防护研究院辐射防护及环境保护专业在读研究生。E-mail:liweiming1997@126.com

Establishment of γ-ray air kerma (radiotheraph lever) standard device

LI Weiming1, WEI Yingjing1,2, LI Yin1, CHEN Shuangqiang1, WANG Mingliang1, HAO Shidong1   

  1. 1. China Institute for Radiation Protection,Taiyuan 030006;
    2. Tsinghua University,Beijing 100084
  • Received:2021-01-14 Online:2021-10-25 Published:2021-11-26

摘要: 采用活度为1.85×1014 Bq的60Co放射源建立了γ射线空气比释动能(治疗水平)标准装置,同时从实际情况和安全性考虑设计建造了一套附属的安全联锁系统。为了评价标准装置的性能指标,依次测试辐射场的空气比释动能率范围、辐射野、均匀性以及散射等。测量结果表明在1~5 m的距离内,在不加铅衰减的条件下辐射场的空气比释动能率范围为1.71~43.5 Gy/h,加衰减器后最小空气比释动能率可达1.5×10-4 Gy/h。在无衰减器的条件下,该标准装置在距离放射源1~5 m范围内平方反比律在1%以内成立;在加衰减器的条件下,该标准装置在距离放射源2~5 m范围内平方反比律在±3.5 %以内成立。2 m位置处空气比释动能率波动在5 %以内的辐射野半径为13 cm,空气比释动能率波动在1 %以内的辐射野半径为7 cm。标准装置的技术性能指标满足开展治疗水平剂量仪的检定/校准要求。

关键词: 60Co γ射线, 安全联锁, 治疗水平, 空气比释动能率

Abstract: A standard device for γ-ray air kerma (radiotherapy) was established using 60Co radioactive source with an activity of 1.85×1014Bq. Meanwhile, a safety interlock system with six subsystems and logic of “and” gate was designed for the standard device. In order to evaluate the performance indicators of the standard device, the air kerma rate range of the radiation field, the radiation field, uniformity and scattering were measured in sequence. The measurement results show that at a distance of 1 m to 5 m, the air kerma rate of the radiation field without lead attenuation ranges from 1.71 Gy/h to 43.5 Gy/h, and the minimum air kerma is after adding the attenuator The kerma rate can reach 1.5×10-4 Gy/h. Under the condition of no attenuator, the inverse square law of the standard device is established within 1% in the range of 1 m to 5 m from the radiation source. Under the condition of adding the attenuator, the inverse square law of the standard device is established within 3.5% in the range of 1 m to 5 m from the radiation source. At a position of 2 m, the radius of the radiation field where the air kerma rate fluctuates within 5% is 13 cm, and the radius of the radiation field where the air kernal energy rate fluctuates within 1% is 7 cm. The technical performance indicators of the standard device meet the verification/calibration requirements of the radiotherapy dosimeter.

Key words: 60Co γ ray, safety interlock system, radiotherapy, air kerma rate

中图分类号: 

  • TL84