辐射防护 ›› 2017, Vol. 37 ›› Issue (3): 169-173.

• 剂量学基础 • 上一篇    下一篇

SOI 硅微剂量探测器对中子和伽马辐射场线能谱测量的GEANT4模拟研究

雷鸣1, 刘书焕2, 宗鹏飞1, 刘兵1   

  1. 1.中国核动力研究设计院,成都 610005;
    2.西安交通大学,西安 710049
  • 收稿日期:2016-07-29 出版日期:2017-05-20 发布日期:2021-11-11
  • 通讯作者: 刘书焕。E-mail: liushuhuan@mail.xjtu.edu.cn
  • 作者简介:雷鸣(1989—),男,2011年毕业于南华大学核工程与核技术专业,2014年毕业于西安交通大学核科学与技术专业,获工学硕士学位,助理工程师。E-mail: minglei2001@163.com
  • 基金资助:
    国家自然科学基金面上项目(11175139)资助。

GEANT4 simulation of silicon-on-insulator microdosimeter for monitoring lineal spectra of neutron and gamma mixed field

Lei Ming1, Liu Shuhuan2, Zong Pengfei1, Liu Bing1   

  1. 1. Nuclear Power Institute of China, Chengdu 610005;
    2. Xi'an Jiaotong University, Xi'an 710049
  • Received:2016-07-29 Online:2017-05-20 Published:2021-11-11

摘要: 绝缘体上硅(SOI)硅微剂量测量系统可以通过测量中子和伽马混合辐射场的线能谱来获取不同类型辐射的剂量当量贡献。本文根据SOI硅微剂量探测器物理设计,采用GEANT4软件建模对Cf-252中子和伽马混合辐射场及Co-60伽马辐射线能谱测量进行蒙特卡罗模拟,并进一步分析了SOI探测器转换层对伽马线能谱测量的影响。结果表明,SOI硅微剂量探测器能够区分中子和伽马的剂量贡献,并且伽马线能谱峰值随着转换层厚度发生变化,有可能利用该特性实现不同贯穿深度下伽马辐射剂量贡献的测量。模拟分析结果可为SOI硅微剂量探测器设计及应用提供参考。

关键词: SOI, 硅微剂量探测器, 线能谱, 剂量当量, GEANT4, 混合场

Abstract: Lineal energy spectra of novel SOI (silicon-on-insulator) microdosimeter and detector response in Cf-252 neutron and gamma mixed field and Co-60 gamma radiation field were simulated with GEANT4. The simulated results were compared with the TEPC monitoring experiments with good consistency. Some differences are identified and discussed. The effects of different conversion layer thickness on lineal energy spectra were analyzed. Through the simulation, it can be seen that the SOI detector could be used to distinguish the contribution of neutron from gamma. We also found that the thicker conversion layer could increase the peak values of gamma lineal spectra, which may be used for gamma dose measurement at various depth of tissue. These results provide valuable information for the design of the SOI microdosimeter.

Key words: SOI, microdosimeter, lineal energy, dose equivalent, GEANT4, mixed fiel

中图分类号: 

  • TL814