辐射防护 ›› 2024, Vol. 44 ›› Issue (5): 454-460.

• 辐射防护监测 • 上一篇    下一篇

β辐射个人剂量当量Hp(3)次级标准电离室的设计与性能测试

李胤1, 王川2, 以恒冠1, 刘新昊1, 王桢1, 段嘉宇1   

  1. 1.中国辐射防护研究院 辐射安全与防护山西省重点实验室,太原 030006;
    2.核电运行研究(上海)有限公司,上海 200126
  • 收稿日期:2023-11-02 出版日期:2024-09-20 发布日期:2024-11-05
  • 通讯作者: 以恒冠。E-mail:kidyhg1412@sina.com
  • 作者简介:李胤(1995—),男,2017年毕业于湖北科技学院核工程与核技术专业,2020年毕业于兰州大学核技术及应用专业,获硕士学位,助理研究员。E-mail:liyin_cirp@163.com

Design and performance test of secondary standard ionization chamber for β radiation personal dose equivalent Hp(3)

LI Yin1, WANG Chuan2, YI Hengguan1, LIU Xinhao1, WANG Zhen1, DUAN Jiayu1   

  1. 1. China Institute of Radiation Protection, Shanxi Provincial Key Laboratory of Radiation Safety and Protection, Taiyuan 030006;
    2. Nuclear Power Operations Research Institute, Shanghai 200126
  • Received:2023-11-02 Online:2024-09-20 Published:2024-11-05

摘要: 研发了β辐射个人剂量当量Hp(3)次级标准电离室,不同于主标准器外推电离室的外推结构设计,该电离室采取附带体模的平板电离室的结构,兼顾高测量精度(相对固有误差<2.0%)的同时,简化试验流程。辐射性能试验结果表明:在β辐射公式p(3)个人剂量当量率5~170 mSv/h内的相对固有误差不超过±1.5%;具备β辐射公式p(3)个人剂量当量率0.5 mSv/h~1.5 Sv/h内至少5个量程的测量范围且相对固有误差不超过±4.0%;在β辐射±60°入射角下的响应相对偏差不超过±5%;电离室的测量重复性为0.6%;对90Sr/90Y源在30 cm处加展平过滤的测量相对扩展不确定度为Urel = 3.9%(k = 2)。另外,实验发现,Hp(3)电离室对低能污染光子具有相对较高的响应。

关键词: β辐射, 个人剂量当量Hp(3)次级标准电离室, 相对固有误差, 角响应, 不确定度

Abstract: In this paper, a secondary standard ionization chamber for β radiation personal dose equivalent Hp(3) is developed. Different from the extrapolation structure design of the primary standard extrapolation ionization chamber, the secondary standard ionization chamber adopts the structure of a flat plate ionization chamber with a phantom, and takes into account high measurement accuracy (relative inherent error <2.0%) while greatly simplifying the complexity of the test process. The radiation performance test results show that relative inherent error within the individual dose equivalent rate of beta radiation Hp(3) from 5 mSv/h to 170 mSv/h does not exceed ±1.5%; with regard to beta radiation Hp(3), for measurement range of at least 5 orders within the personal dose equivalent rate of 0.5 mSv/h-1.5 Sv/h, the relative inherent error does not exceed ±4.0%; the relative deviation of the response under the incident angle of β radiation ±60° does not exceed ±5%; measurement repeatability of the ionization chamber is 0.6%; relative expanded uncertainty of the measurement with flattening filter at 30 cm for the 90Sr/90Y source is Urel=3.9% (k=2). In addition, it was found experimentally that the Hp(3) ionization chamber has a relatively high response to low-energy contamination photons.

Key words: beta radiation, secondary standard ionization chamber for personal dose equivalent Hp(3), relative inherent error, angular response, uncertainty