Application of Portable Hydrogen Meter in Nuclear Power Plant

Chen Zhanggui, Bai Zhaotao, Wu Huaqiang, Zhan Xiaocong, Wang Long, Zhang Tao   

  1. Fujian Ningde Nuclear Power Co., Ltd., Fujian Fudin, 355200
  • Received:2014-08-15 Online:2014-10-20 Published:2025-02-24

Abstract: During nuclear power plant operation and overhaul, the hydrogen concentration in reactor coolant system is subject to monitoring. This paper mainly describes the experimental process of rapid measurement of hydrogen in reactor coolant system by using portable ORBISPHERE 3654 hydrogen meter and the application of portable hydrogen meter in Ningde nuclear power plant. The results show that the portable hydrogen meter can rapidly and accurately measure hydrogen content in the RCP with optimal water sample flow at a rate of 270-310 mL/min.