RADIATION PROTECTION BULLETIN ›› 2014, Vol. 34 ›› Issue (5): 44-46.
Previous Articles
Chen Zhanggui, Bai Zhaotao, Wu Huaqiang, Zhan Xiaocong, Wang Long, Zhang Tao
Received:
Online:
Published:
Abstract: During nuclear power plant operation and overhaul, the hydrogen concentration in reactor coolant system is subject to monitoring. This paper mainly describes the experimental process of rapid measurement of hydrogen in reactor coolant system by using portable ORBISPHERE 3654 hydrogen meter and the application of portable hydrogen meter in Ningde nuclear power plant. The results show that the portable hydrogen meter can rapidly and accurately measure hydrogen content in the RCP with optimal water sample flow at a rate of 270-310 mL/min.
Key words: Reactor coolant system, Portable hydrogen meter, Hydrogen concentration  ,  , (
CLC Number:
Chen Zhanggui, Bai Zhaotao, Wu Huaqiang, Zhan Xiaocong, Wang Long, Zhang Tao. Application of Portable Hydrogen Meter in Nuclear Power Plant[J].RADIATION PROTECTION BULLETIN, 2014, 34(5): 44-46.
0 / / Recommend
Add to citation manager EndNote|Reference Manager|ProCite|BibTeX|RefWorks
URL: http://journal01.magtech.org.cn/Jwk3_fsfhtx/EN/
http://journal01.magtech.org.cn/Jwk3_fsfhtx/EN/Y2014/V34/I5/44
Cited