RADIATION PROTECTION ›› 2024, Vol. 44 ›› Issue (5): 461-468.

Previous Articles     Next Articles

Correction of crosstalk influence in the sample measurement of total α/β used by ultra low background flow gas proportional counter

XU Hailiang1, CHEN Cheng1, WANG Liang1, ZHU Zhu2, GUO Xuying1, WANG Qian1, SU Kunpu1   

  1. 1. Sichuan Radiation Environment Management and Monitoring Center Station,Chengdu 611100;
    2. Key Laboratory of Mass Spectrometry Analysis Technology of China National Nuclear Corporation,Chengdu 611130
  • Received:2023-08-14 Online:2024-09-20 Published:2024-11-05

Abstract: The ultra-low background flow gas proportional counter is the ideal detector for total α/β sample measurement, but itself has strong crosstalk phenomenon, which affects the accuracy of sample measurement. This paper systematically studies the flow type proportional counter under the same mass thickness, different 241Am and 40K mass fractions α→β, β→α Cross channel influence. This paper established a binary equation for total α/β channel correction and compared it with the cover correction method and with the self-made 25 groups α/β simulate powder samples for measurement and channel correction. The results show that: The impact of α→β crosstalk is significant; The change with α counting rate is not constant; The influence of the β→α channel correction factor is relatively small, with an average value of 0.001. The accuracy of the measurement results is poor by using the cover correction method to correct the simulated powder β sample, with absolute relative deviation ranging from 1.4% to 45.0%. The binary equation correction method is used to correct the simulated powder β sample. The accuracy of the measurement results is relatively high, with an absolute value of relative deviation ranging from 0 to 5.5%.

Key words: ultra low background flow proportional counter, cross-talk, binary equation correction method

CLC Number: 

  • TL811+.2