RADIATION PROTECTION ›› 2024, Vol. 44 ›› Issue (1): 57-61.

Previous Articles     Next Articles

Comparison and analysis of electromagnetic radiation monitoring instruments in anechoic chamber

NING Jian1, LI Zhanyou1, ZHOU Feng2, YUAN Xiuhua2   

  1. 1. Environmental Radiation Monitoring and Nuclear Emergency Response Technical Support Center of Guangdong Province, Guangzhou 510300;
    2. China Academy of Information and Communications Technology,Beijing 100191
  • Received:2023-08-22 Online:2024-01-20 Published:2024-02-05

Abstract: In order to improve the comparability of the comparison results of monitoring instruments, it is necessary to select a place where the field strength is stable and is not affected by the external electromagnetic environment for comparison. This article analyzes and summarizes the experience and practices of conducting frequency-selective radio frequency electromagnetic radiation monitoring instrument comparison-measurement in a standard anechoic chamber, and finds out some problems that are difficult to detect in daily work.This paper also proposes suggestions for monitoring instrument verification during the monitoring period.Conducting comparisons in a stable and standard electromagnetic field environment could provide a new practice for electromagnetic environment monitoring instrument comparison.

Key words: frequency-selecting, 5G NR signal, electromagnetic radiation monitoring instrument, anechoic chamber, comparative measurement

CLC Number: 

  • X85