RADIATION PROTECTION ›› 2024, Vol. 44 ›› Issue (1): 33-41.

Previous Articles     Next Articles

A performance study of a new personal neutron dosimeter based on SSNTD

JIAO Yan1,2, CAO Qinjian1,2, WEI Xiaofeng1,2, XU Yuan1,2, ZHAO Yuan1,2, LIU Liye1,2   

  1. 1. China Institute for Radiation Protection, Taiyuan 030006;
    2. Shanxi Key Laboratory for Radiation Safety and Protection, Taiyuan 030006
  • Received:2022-08-30 Online:2024-01-20 Published:2024-02-05

Abstract: As a passive personal neutron dosimeter, CR-39 solid state nuclear track detector is convenient, cheap, insensitive to gamma and beta rays and has strong anti-interference ability, and better response to fast neutrons. A new personal neutron dosimeter has been developed combining with two measurement methods of CR-39 and albedo thermoluminescence dosimeter. Performance experiments are carried out on the CR-39 detector according to the international standard ISO 21909-1. The results show that the CR-39 solid state nuclear track detector has good repeatability, batch homogeneity, dose linearity, stability, response of reference radiation field and no photon-sensitivity. The detection threshold and uncertainty of dose measurement are presented in this paper. It provides an experimental data reference for the introduction of international standard.

Key words: external exposure, neutron, personal dosimeter, CR-39

CLC Number: 

  • TL818+.1