RADIATION PROTECTION ›› 2021, Vol. 41 ›› Issue (5): 415-421.

Previous Articles     Next Articles

The discussion on gamma-rays characteristic peak interference correction method when analyzing radionuclide activity with gamma spectrometry

YANG Xiuyu1, MENG Jiqun1, LI Qingguang1, WEI Yingjing2   

  1. 1. 404 Company Limited, CNNC,Lanzhou 732850;
    2. China Institute for Radiation Protection,Taiyuan 030006
  • Received:2020-07-22 Online:2021-09-20 Published:2021-11-26

Abstract: Aiming at the problem of gamma-ray characteristic peak interference correction in gamma spectrometry during a CNNC Environmental monitoring labs comparison, this paper conducts measurement experiment of uranium, thorium, radium and potassium in soil. When calibrating the efficiency of the spectrometer with natural soil standard source, it is necessary to analyze whether the gamma-ray characteristic peaks are interfered by other rays, and correct the efficiency by modifying the Nuclide activity value when calculating efficiency. When analyzing the nuclide activity of the samples by Gamma-Vision software, the activity correction should be carried out if the nuclide activity calculated by different characteristic rays is quite different. The characteristic peaks (such as 235U 185.7 keV, 238U 92.6 keV) used in the calculation of nuclide activity are analyzed, the contribution of the interference peak to the measured energy spectrum (Overlapping Peak) activity is deducted. This method gives qualified 238U,226Ra,232Th,40K and 137Cs data in CNNC soil sample comparison.

Key words: gamma spectrometry, gamma-ray characteristic peak interference correct, Gamma-Vision software analysis method