RADIATION PROTECTION ›› 2020, Vol. 40 ›› Issue (6): 561-564.

• Radiation Detection Technology and Application • Previous Articles     Next Articles

Main performance test of large area gas-sealed proportional counter

QIAO Li1,2, RAO Xianming3, DU Xiangyang1,2, REN Yi1,2, ZHANG Jia1,2, GUO Xirong1,2, WANG Yanfei2   

  1. 1. China Institute for Radiation Protection,Taiyuan 030006;
    2. Shanxi Zhongfu Nuclear Instruments Co. Ltd.,Taiyuan 030006;
    3. China Nuclear Xiapu Nuclear Power Co. Ltd.,Fujian Ningde 355100
  • Online:2020-11-20 Published:2021-01-27

Abstract: Shanxi Zhongfu Nuclear Instruments Co.,Ltd.(CIRNIC) has independently developed and produced various types of large area gas-sealed proportional counter of 40-600 cm2.An experiment focusing on background count rate,plateau curve,surface emission rate response,coefficient of variation and other physical characters,on one of the device of which with 517 cm2 effective area has shown that under the same testing environment and experimental conditions,background count rate is 24.5 cps,surface emission rate 50.4%,coefficient of variation 0.85%.

Key words: gas-sealed proportional counter, plateau curve, surface emission rate response, coefficient of variation

CLC Number: 

  • TL811+.2