RADIATION PROTECTION ›› 2020, Vol. 40 ›› Issue (6): 550-555.

• Radiation Detection Technology and Application • Previous Articles     Next Articles

Study on reducing the low detection limit of net count rate by low backgroundα, β measurement instrument from theoretical point of view

JIN Tao, SUN Yu, ZHANG Zhipeng, WU Yao, SONG Jigao, LI Junjie, LUO Yuanpan, PEI Min, ZENG Bo   

  1. Nuclear Power Institute of China,Sichuan Provincial Engineering Laboratory of Nuclear FacilityDecommissioning and Radwaste Management,Sichuan Chengdu 610065
  • Online:2020-11-20 Published:2021-01-27

Abstract: In this paper,the formula of the low detection limit of the net count rate in radioactive measurement is derived from the point of non-equality,the actually probability of making type II errors corresponding to the two typical low detection limit is calculated.In the absence of an approximation,the number of radionuclides that have decayed follow a binomial distribution.When the average of the net count rate is the low detection limit of the net count rate,the probability that the random value of the net count rate is less than the decision threshold is the probability value of making type II errors.Combining the number of radioactive nuclei at initial moment,the number of radioactive nuclei in the whole measurement time that has decayed and the probability of radioactive nuclides decay,the actual probability value of making type II errors can be calculated.It is found that the maximum practical probability value of the type II errors is 1 to 2 orders of magnitude less than the nominal value in most case.Therefore,the low detection limit can be modified,and an appropriate correction coefficient can be given,so that the actual probability value of making type II errors is close to and slightly less than the nominal value.Through calculation,it is found that the correction coefficient given by kα=1.645 and kβ=1.645 can reduce low detection limit by 22 percent.

Key words: net count rate, low detection limit, correction coefficient

CLC Number: 

  • TL817