Factors affecting detection limits of aerosol sampling in Fangchenggang nuclear power plant

HOU Jie1, LI Wenting1, HUANG Li1, HAN Shanbiao1, GAO Yuzhong2, WANG Ruiying1, LI Li1, LI Bing1, YUE Huiguo, CHEN Xiaoqiu1   

  1. 1. Nuclear and Radiation Safety Center, Ministry of Ecology and Enviroment, Beijing 100082;
    2. Fangchenggang Nuclear Power Station Co., Ltd., Guangxi Fangchenggang 538001
  • Received:2018-03-09 Online:2018-11-20 Published:2020-07-21

Abstract: The low limit detection (LLD) and the influencing factors for air effluent sampling monitoring in first stage of Fangchenggang nuclear power plant was studied. The LLD of different radioactive isotopes in air effluent with certain relative standard deviation under certain confidence level was calculated. The relationship of the relative standard deviation varying with net counting rate in different counting time was analyzed. The results showed that only when the radioactive release of air effluent approached to the calculated level, the air effluent monitoring was able to get a meaningful outcome. The analysis provides a reference for other nuclear power stations to make their own air effluent monitoring program.

Key words: Fangchenggang NPP, air effluent, low limit detection, aerosol sampling system