RADIATION PROTECTION ›› 2018, Vol. 38 ›› Issue (1): 45-50.
Previous Articles Next Articles
CHANG Yu, MENG Jinhong, ZHONG Xueming, ZHANG Qingjie
Received:
Online:
Published:
Abstract: Measuring device for tritium in process exhaust using ionization chamber (effective volume 11.0 mL) was developed after material selection, structure design and parameter optimization. The calibration curve was plotted with a good linearity in the range of 0.013 4%—0.168% (v). The results showed that the detection lower limit of the device was 1.35×10-4 %(v); the coefficient of variation was no more than 1.50%; and the response time was less than 10 s. The process measurement requirements were fairly met.
Key words: ionization chamber, tritium, background current, memory effect, calibration curve
CLC Number:
CHANG Yu, MENG Jinhong, ZHONG Xueming, ZHANG Qingjie. Development of a measuring device for tritium in process[J].RADIATION PROTECTION, 2018, 38(1): 45-50.