RADIATION PROTECTION ›› 2014, Vol. 34 ›› Issue (5): 330-332.

Previous Articles     Next Articles

Intercomparison of External Dose Measurements by Eight Entities of Jiangsu Province

Wang Wanping, Zhang Ruixue, Shen Leyuan, Gao Xiaofeng, Zhang Haiqian   

  1. Jiangsu Radiation Environment Management and Monitoring Center, Nanjing 210019
  • Received:2013-12-31 Online:2014-09-20 Published:2025-01-06

Abstract: In Jiangsu province, eight agencies participated in the intercomparison of external dose measurement. In comparison of 160 thermoluminescent dosimeter, the results showed that only one group of blind sample 1 was abnormal. The relative deviation of other 7 groups compared to the standard dose of 1 mSv was -16%—1%. The relative deviation of blind sample 2 compared to the standard dose of 4 mSv was -18.5% to 3.25%.The results showed that the measurement of seven laboratories were satisfactory.

Key words: TLD, dosimeters, cumulative dose, intercomparison

CLC Number: 

  • R144.1