Design of a Detector-separated Neutron Probe

Liang Jing, Feng Xiaobo, An Yanlong, Liu Jianzhong   

  1. Naval Armament Department, Taiyuan, Shanxi, 030000;
    China Institute for Radiation Protection, Taiyuan, Shanxi, 030006
  • Received:2021-12-28 Online:2022-04-20 Published:2022-07-21

Abstract: We have designed a neutron measurement system with the preamplifier separated from the detector. The electronic circuit with poor irradiation resistance from the detector assembly is isolated, and connected remotely through a coaxial cable. The test results show that, the maximum relative inherent error of the probe is only 8.6%, and the difference between the relative response of the Am-Be source and the 252Cf source is 9%. Its performance meets the design requirements.

Key words: Neutron measurement system, Detector assembly, Electronic circuit, Separated