辐射防护通讯 ›› 2026, Vol. 46 ›› Issue (4): 23-28.

• 研究通报 • 上一篇    下一篇

常见α/β表面污染监测仪对14C β射线的表面发射率响应分析

李小双, 陆小军, 宋家斑   

  1. 上海市计量测试技术研究院有限公司,上海 201203
  • 收稿日期:2026-04-16 出版日期:2026-08-20 发布日期:2026-09-07
  • 通讯作者: 陆小军。E-mail:luxj@simt.com.cn
  • 作者简介:李小双(1990—),男,2013年毕业于中国辐射防护研究院辐射防护及核安全专业,获硕士学位,现从事电离辐射计量工作,E-mail:2386836410@qq.com
  • 基金资助:
    国家市场监督管理总局科技项目(No.2024MK023)。

Analysis of the surface emission rate response of common α/β surface contaminationmonitors to 14C β rays

LI Xiaoshuang, LU Xiaojun, SONG Jiaban   

  1. Shanghai Institute of Measurement and Testing Technology, Shanghai 201203
  • Received:2026-04-16 Online:2026-08-20 Published:2026-09-07

摘要: 为明确不同类型α/β表面污染监测仪对低能β射线的探测性能差异及影响机制,本文以14C平面源(β射线最大能量156 keV)为典型研究对象,选取配备闪烁体探测器、盖革-米勒(G-M)计数器及正比计数器的多款商用表面污染监测仪,系统开展表面发射率响应特性实验研究。结果表明:不同类型仪器对14C低能β射线的响应表现不同,其中闪烁体探测器响应范围跨度较大,为0.2%~21.1%,盖革-米勒计数器响应范围为8.3%~12.2%,正比计数器响应为7.9%;相同材质探窗,质量厚度与仪器对14C β射线的响应值呈负相关,穿透率与响应值呈正相关。研究结果可为14C的表面污染监测提供参考,在表面污染仪国产化研发、性能优化及用户选型方面也具有一定的价值。

关键词: 低能β, 14C, 表面发射率响应, 表面污染测量, 探测窗, 穿透率

Abstract: To investigate the differences in detection performance and the underlying influence mechanisms of various types of α/β surface contamination monitors for low-energy beta rays, this paper selects a 14C planar source (with a maximum beta energy of 156 keV) as a typical research subject and systematically conducts an experimental study on surface emission response characteristics using multiple commercial monitors equipped with scintillation detectors, Geiger-Müller (G-M) counters, and proportional counters. The results indicate that different instrument types exhibit varying responses to low-energy beta rays from 14C. Specifically, the scintillation detectors show a wide response range from 0.2% to 21.1%, the G-M counters show a response range of 8.3% to 12.2%, and the proportional counter yields a response of 7.9%. However, for windows made of the same material, the mass thickness is negatively correlated with the instrument′s response to 14C beta rays, while the penetration rate, which is determined by the mass thickness, is positively correlated with the response value. These findings can provide important technical references for the surface contamination monitoring of 14C, and also hold practical value for the domestic development of surface contamination monitors, performance optimization, and user equipment selection.

Key words: low-energy β, 14C, surface emission rate response, surface contamination measurement, detection window, penetration rate

中图分类号: 

  • TL84