辐射防护通讯 ›› 2026, Vol. 46 ›› Issue (1): 1-5.

• 研究通报 • 上一篇    下一篇

高纯3He气体中氚分析装置研制及性能测试

陈伟斌1,2,3, 朱波1,2, 陈俊良1, 王迪4, 5, 陆宗燕1,2, 马治军1,3   

  1. 1.中核四0四有限公司,兰州 732850;
    2.中核特种材料提取存储技术重点实验室,兰州 732850;
    3.兰州大学核科学与技术学院,兰州 730000;
    4.山西中辐核仪器有限责任公司,太原 030006;
    5.中国辐射防护研究院,太原 030006
  • 收稿日期:2025-12-09 出版日期:2026-02-20 发布日期:2026-03-05
  • 通讯作者: 朱波。E-mail:253322381@qq.com
  • 作者简介:陈伟斌(1994—),男,2021年毕业于哈尔滨工程大学核能与核技术工程专业,获硕士学位,工程师。E-mail:chenweibin2013@163.com
  • 基金资助:
    中核集团“青年英才”资助项目([2024]310号)。

Development and performance testing of a tritium analysis device in high-purity 3He gas

CHEN Weibin1,2,3, ZHU Bo1,2, CHEN Junliang1, WANG Di4, 5, LU Zongyan1,2, MA Zhijun1,3   

  1. 1. The 404 Company Limited., CNNC, Lanzhou 732850;
    2. CNNC Key Labortory on Special Material Extraction and Storage Technology, Lanzhou 732850;
    3. School of Nuclear Science and Technology, Lanzou University, Lanzhou 730000;
    4. Shanxi Zhongfu Nuclear Instruments Co., Ltd., Taiyuan 030006;
    5. China Institute for Radiation Protection, Taiyuan 030006
  • Received:2025-12-09 Online:2026-02-20 Published:2026-03-05

摘要: 高纯3He气体产品中的氚活度浓度测量工作,对分析装置的探测下限(MDA)及样品损耗量提出了严格要求。本研究针对高纯3He气体中氚活度浓度的测量需求,以正比计数器为探测元件,研制出一套分析装置,并系统地评估了该装置坪特性、探测效率及探测下限等关键参数。结果表明:该装置的坪长为150 V,坪斜为4.31%/100V;测得氚信号能谱与本底信号能谱表现出明显差异;对氚的探测效率可达66.78%;探测下限为2.55 Bq/L,且单次测量仅需消耗 2 mL 样品。该装置性能可靠,能够为高纯3He气体中氚的准确测量提供有效的技术支撑。

关键词: 3He气体, 氚分析装置, 正比计数器, 探测下限

Abstract: The measurement of tritium activity concentration in high-purity 3He gas products imposes stringent requirements on the detection limit and sample consumption of analytical devices. To address the requirements for measuring tritium activity concentration in high-purity 3He gas, this study developed an analytical device using a proportional counter as the detection component through structural design and material optimization. The key parameters of the device, such as plateau characteristics, detection efficiency, and detection limit, were systematically evaluated. The results show that the plateau length of the device is 150 V with a plateau slope of 4.31% per 100 V. The measured tritium signal spectrum shows significant differences from the background signal spectrum. The detection efficiency for tritium reaches 66.78%, and the detection limit is 2.55 Bq/L, with a sample consumption of only 2 mL per measurement. The device demonstrates reliable performance and provides effective technical support for accurate tritium measurement in high-purity 3He gas.

Key words: 3He gas, tritium analysis device, proportional counter, detection limit