辐射防护通讯 ›› 2016, Vol. 36 ›› Issue (1): 1-9.

• 专题报告 • 上一篇    下一篇

全国辐射环境监测网络环境γ辐射剂量率测量比对

吴昱城, 胡丹, 赵顺平, 程建平1, 宋建锋, 曾志1, 丁逊, 徐斌2, 胡飞, 许宏   

  1. 环境保护部辐射环境监测技术中心,国家环境保护辐射环境监测重点实验室, 杭州, 310012;
    1.清华大学,粒子技术与辐射成像教育部重点实验室, 北京, 100084;
    2.四川省辐射环境管理监测中心站, 成都, 610031
  • 收稿日期:2015-11-28 出版日期:2016-02-20 发布日期:2021-11-19
  • 作者简介:吴昱城(1987- ),男, 2008年毕业于清华大学工程物理专业,学士; 2013年毕业于清华大学核科学与技术专业,博士; 高级工程师。

Intercomparison of Environmental Gamma Radiation Dose Rate Measurements within the National Radiation Environmental Monitoring Network of China

Wu Yucheng, Hu Dan, Zhao Shunping, Cheng Jianping1, Song Jianfeng, Zeng Zhi1, Ding Xun, Xu Bin2, Hu Fei, Xu Hong   

  1. Radiation Monitoring Technical Center of Ministry of Environment Protection, State Key Laboratory of Radiation Environmental Monitoring, MEP, Hangzhou, 310012;
    1. Department of Engineering Physics, Tsinghua University, Key Laboratory of Particle and Radiation Imaging of Ministry of Education, Beijing, 100084;
    2. Sichuan Management & Monitoring Center Station of Radioactive Environment,Chengdu, 610031
  • Received:2015-11-28 Online:2016-02-20 Published:2021-11-19

摘要: 2013年12月,环境保护部辐射环境监测技术中心在四川省凉山彝族自治州组织了全国辐射监测网络的环境γ辐射剂量率和宇宙射线响应测量比对。共有32家成员单位的47台仪器设备参与了7个环境测点的比对测量,比对结果统一到γ空气吸收剂量率,采用Z比分数值分别对不同型号仪器的测量结果进行评价。比对结果显示,闪烁计数器测量数据的离散性较高压电离室大。除宇宙射线响应测点外,两类仪器的测量结果保持着较好的线性。

关键词: γ辐射剂量率, 剂量率仪, 比对, 刻度

Abstract: In December 2013, an intercomparison of environmental gamma radiation dose rate and cosmic ray response measurements within the national radiation environmental monitoring network was held in Liangshan Yi Autonomous Prefecture, Sichuan Province, which was organized by RadiationMonitoring Technical Center of Ministry of Environment Protection. A total of 47 sets of instruments from 32 member units participated in the intercomparison measurements at 7 environmental measuring points. All the results are unified interms of absorbed dose rate in air. Results of different types of instruments are independently evaluated by Z score values. It is found that the data of scintillation counters have relatively wider dispersion than those of high pressure ionization chambers. Besides the cosmic ray response points, the results of both keep good linearity.

Key words: γ dose rate, Dose ratemeter, Intercomparison, Calibration

中图分类号: 

  • X837