Direct measurement of gross α/β in wastewater by liquid scintillation method

WANG Ruijun1, BAO Li 1, CHEN Ling2, REN Xiaona1, LIU Wangdong1, SHI Qiqi1, CUI Linhong1   

  1. 1. China Institute for Radiation Protection, Taiyuan 030006;
    2. China Academy of Atomic Energy, Beijing 102413
  • Received:2024-02-27 Online:2025-09-20 Published:2026-01-14

Abstract: When gross α/β in wastewater is high, liquid scintillation spectrometer with α/β discrimination function can be used to directly determine gross α/β, which has the outstanding advantages of simple process, low background and high counting efficiency, and the result can be analyze by considering the liquid scintillation count spectrum. In this paper, the concentration of gross α/β activity in wastewater was directly determined by LSC, and the results were compared and analyzed. The main contributing nuclide of gross β in this batch of wastewater samples is 90Sr, and the activity concentration of gross β is three orders of magnitude higher than that of gross α. The relative deviations of gross α and gross β measured by liquid scintillation method and gas-flow proportional counting method are 2.2%-22.5% and 2.3%-16.7%, respectively, and the correlation is good.

Key words: LSC, gross α, gross β, wastewater