RADIATION PROTECTION ›› 2024, Vol. 44 ›› Issue (6): 612-622.

Previous Articles     Next Articles

Influence and correction of sample mass thickness variation for total β radioactivity measurement in water

DU Yunwu, DENG Xiaoqin, BI Chaowen, ZHU Jie, WANG Liang, WANG Qian, LV Yingfu, WANG Yangyang, HE Dai   

  1. Sichuan Provincial Radiation Environment Management and Monitoring Center Station, Chengdu 610031
  • Received:2023-09-18 Published:2025-01-06

Abstract: By using two sample media, the detection efficiency of natural and artificial β nuclides with different mass thickness was simulated by experiment and Monte Carlo Method. The relationship between detection efficiency, count rate, detection efficiency correction factor, reciprocal of detection efficiency and mass thickness, as well as detection efficiency and energy were analyzed. The results show that 40K is used as the standard source, the detection efficiency correction factor are needed to correct the detection efficiency. The measurement of β radioactivity of 234Th, 210Pb, 212Pb, 228Ra, 60Co, 99Tc, etc. should be modified by thin source method and detection efficiency correction factor. The reciprocal of detection efficiency is linear with the mass thickness, which can be used to scale the detection efficiency better. The detection efficiency is exponentially related to the average energy of beta rays. The β ray equivalence of KCl and CaSO4 media is good and can meet the requirements of relevant standards.

Key words: β nuclide, mass thickness, detection efficiency, correction factor

CLC Number: 

  • X830