RADIATION PROTECTION BULLETIN ›› 2014, Vol. 34 ›› Issue (3): 28-31.

• Experiences • Previous Articles     Next Articles

Comparison of Environmental Gamma Radiation Dose Rates in Zhejiang Province in 2012

He Bisheng, Ding Xun, Shang Jiangfeng, Liu Gongye, Xu Hong   

  1. Zhejiang Province Environmental Radiation Monitoring Center, Hangzhou, 310012
  • Received:2013-08-29 Online:2014-06-20 Published:2025-02-26

Abstract: A comparison of environmental radiation dose rate in Zhejiang province was performed by Zhejiang Radiation Environmental Monitoring Station in 2012 in order to evaluate the quality of radiation environmental monitoring. The robust statistical Z score method was taken to evaluate the consistency of measuring results in this work. The monitoring resultsfrom 20 instruments at 5 comparison points showed that outlier, questionable value and satisfactory value was 5%, 6% and 89% of the total, respectively. We found the problems of instruments through this comparison to ensure the accuracy ofmonitoring results.

Key words: Environmental, Gamma radiation dose rate, Comparison of Measurement

CLC Number: 

  • X837