RADIATION PROTECTION BULLETIN ›› 2018, Vol. 38 ›› Issue (4): 37-40.
Previous Articles Next Articles
Feng Yingsi, Jiang Lan, Zhang Yanjin, Deng Fei
Received:
Online:
Published:
Abstract: The influence of air intake system, scintillation chamber calibration, measuring time and other factors on 226Ra measurement in water by the emanation scintillation method is analyzed. Suggestions for optimization are further proposed as follows: simplifying the intake system and flushing the scintillation chamber and gas path with pure nitrogen; prolonging the sealing time of the standard source, so that the cumulative factor is greater than 0.9, which can improve the stability of the K value of the scintillation chamber; increasing the sample size and measuring time to improve the measurement accuracy, for samples with low radium content.
Key words: 226Ra, Emanation scintillation, Water, Scintillation chamber
CLC Number:
Feng Yingsi, Jiang Lan, Zhang Yanjin, Deng Fei. Some Suggestions for Measuring 226Ra in Water by Emanation Scintillation Method[J].RADIATION PROTECTION BULLETIN, 2018, 38(4): 37-40.