›› 2017, Vol. 43 ›› Issue (4): 57-62.doi: 10.3969/j.issn.1674-1579.2017.04.010
Previous Articles Next Articles
Online:
Published:
Abstract: In order to solve the problem of high precision distance and angle measurement of lidar with large measurement range, a twostep calibration method of lidar sensor based on calibration field is proposed in this paper. Based on the analysis of the angle measurement error and the distance measurement error of lidar, an error correction model is proposed in this paper. This model separates the distance correction from the angle correction. First, the calibration field is used to complete the angle calibration, and the external parameters are also calculated in this step. Second, the baseline field is used to complete the distance calibration. This method reduces the correlation between the distance calibration parameters and the angle calibration parameters, and guarantees the calibration precision and accuracy. The experimental results show that the method is reasonable and effective, and is able to complete the high precision calibration of lidar.
Key words: lidar, twostep calibration, calibration field
CLC Number:
WU Fen-Zhi, GUO Shao-Gang, ZHU Fei-Hu, WANG Li, WU Yun, LIU Da. TwoStep Calibration Method of Lidar Sensor Based on Calibration Field[J]., 2017, 43(4): 57-62.
Add to citation manager EndNote|Reference Manager|ProCite|BibTeX|RefWorks
URL: https://journal01.magtech.org.cn/Jwk3_kjkzjs/EN/10.3969/j.issn.1674-1579.2017.04.010
https://journal01.magtech.org.cn/Jwk3_kjkzjs/EN/Y2017/V43/I4/57
Cited
Analysis on Midcourse Correction of Translunat Trajectory for CE-1